临床和实验室研究年鉴

  • 国际标准期刊号: 2386-5180
  • 期刊 h 指数: 17
  • 期刊引用分数: 6.26
  • 期刊影响因子: 5.31
索引于
  • Genamics 期刊搜索
  • 中国知网(CNKI)
  • 引用因子
  • 研究期刊索引目录 (DRJI)
  • 普布隆斯
  • 欧洲酒吧
  • 谷歌学术
  • 夏尔巴罗密欧
  • 秘密搜索引擎实验室
分享此页面

抽象的

Imprecision Evaluation of Self-Monitoring of Blood Cholesterol (SMBC)Handheld Point of Care Testing Devices: Elemark™ and Cardiochek PA

John Bolodeoku and S Pinkney

The aim of this study was to determine the imprecision of two cholesterol point of care testing devices, the very established CardioChek PA and recently the Elemark™ which is a mobile blood testing device capable of testing multiple parameters for the management of hypercholesterolaemia undergoing certification. The imprecision study of the Elemark™ and CardioChek PA were performed using blood donations from 3 healthy normal lipidemic subjects. There were 10 repetitions for the within run imprecision and the number of measurement days was 6 for between day imprecision. This study showed comparable within run and between day imprecision between the CardioChek PA and the Elemark™. Based on the imprecision guidelines from the NCEP and CRMLN both devices showed reasonable within run imprecision for their total cholesterol and HDLcholesterol estimations but not the LDL-cholesterol calculations. The between day imprecision for both devices was reasonably consistent with the published within person biological variability. With more accuracy and imprecision studies being carried out in understanding these cholesterol testing point of care devices, their value, or lack of, in the lipid monitoring would become more apparent.

免责声明: 此摘要通过人工智能工具翻译,尚未经过审核或验证